UTM For Calibration
Description
Standard : IS 2720 (PART-V) ASTM D 427, BS 1377
Model No. : PIT-046
UTM microcomputer control superimposed force standard machine (pulling, pressing bidirectional) adopts cylinder upward type (the force application point is on the top of the whole machine), four-column structure, high frame rigidity, cylinder, piston drive the upper pressure plate stepless adjustment, pressure is applied hydraulically, using servo pump technology, multi-channel closed-loop coordinated loading of electro-hydraulic Servo control system, microcomputer control coordination multi-stage hydraulic loading, continuous loading and smooth, multi-stage test force maintenance, automatic continuous and stable loading, automatic loading, automatic collection and storage of data, drawing curves, automatic printing of test reports, computer timely control of the test process, display of test force and test curve, simple and reliable operation, easy to operate.
Technical Specification
- UTM, Max capacity 2000KN, Standard load cell calibrator.
- Working area 290×110 (for compression machine) with digital load indicator.
- Max capacity 1000KN, Standard load cell calibrator (for Tension machine) digital load indicator.
Technical Parameters
- Maximum test force: 2000kN
- Test force measurement range: 10%-100% F.S;
- Force error (advance, return): In the range of 10%-100% F.S, the test data were better than ±0.1%.
- Force value repetition: 10%-100% F.S range; The trial data were better than 0.1%.
- Initial load discriminating force: 10%FS, better than 0.03%.
- Negative load volatility: The test data in the range of 10%-100% F.S were better than 0.03%
- Azimuth error: In the range of 10%-100% F.S, the test data is better than 0.2%.
- Horizontality of base and compression table: ≤0.2mm/1000mm
- Parallelism of the upper and lower pressure plates: ≤0 3mm/1000mm
- Flow rate: 0-33 00ml/min
- Control pressure range: 0-25Mpa
- Force holding time: >2h